Characterization and calibration of a triple-GEM detector for medical applications

This project aimed to characterize and calibrate the triple Gas Electron Multiplier (triple-GEM) detector owned by the CRYOMAG research group at Universidad Nacional de Colombia, as well as to determine its possible applications in medical dosimetry. The detector's response to various sources o...

Full description

Autores:
Velásquez Moros, Andrea
Tipo de recurso:
Fecha de publicación:
2019
Institución:
Universidad Nacional de Colombia
Repositorio:
Universidad Nacional de Colombia
Idioma:
spa
OAI Identifier:
oai:repositorio.unal.edu.co:unal/77337
Acceso en línea:
https://repositorio.unal.edu.co/handle/unal/77337
http://bdigital.unal.edu.co/75009/
Palabra clave:
Particle detectors
Detector characterization
Cosmic radiation
Radioactivity
X-rays
Ionization
Radiation dosimetry
Detectores de partículas
Caracterización de detectores
Radiación cósmica
Radiactividad
Rayos X
Ionización
Dosimetría de radiaciones
Rights
openAccess
License
Atribución-NoComercial 4.0 Internacional
id UNACIONAL2_a6d05612add7fb7905cd54fe26ad78ba
oai_identifier_str oai:repositorio.unal.edu.co:unal/77337
network_acronym_str UNACIONAL2
network_name_str Universidad Nacional de Colombia
repository_id_str
dc.title.spa.fl_str_mv Characterization and calibration of a triple-GEM detector for medical applications
title Characterization and calibration of a triple-GEM detector for medical applications
spellingShingle Characterization and calibration of a triple-GEM detector for medical applications
Particle detectors
Detector characterization
Cosmic radiation
Radioactivity
X-rays
Ionization
Radiation dosimetry
Detectores de partículas
Caracterización de detectores
Radiación cósmica
Radiactividad
Rayos X
Ionización
Dosimetría de radiaciones
title_short Characterization and calibration of a triple-GEM detector for medical applications
title_full Characterization and calibration of a triple-GEM detector for medical applications
title_fullStr Characterization and calibration of a triple-GEM detector for medical applications
title_full_unstemmed Characterization and calibration of a triple-GEM detector for medical applications
title_sort Characterization and calibration of a triple-GEM detector for medical applications
dc.creator.fl_str_mv Velásquez Moros, Andrea
dc.contributor.author.spa.fl_str_mv Velásquez Moros, Andrea
dc.contributor.spa.fl_str_mv Castro Serrato, Héctor Fabio
dc.subject.proposal.spa.fl_str_mv Particle detectors
Detector characterization
Cosmic radiation
Radioactivity
X-rays
Ionization
Radiation dosimetry
Detectores de partículas
Caracterización de detectores
Radiación cósmica
Radiactividad
Rayos X
Ionización
Dosimetría de radiaciones
topic Particle detectors
Detector characterization
Cosmic radiation
Radioactivity
X-rays
Ionization
Radiation dosimetry
Detectores de partículas
Caracterización de detectores
Radiación cósmica
Radiactividad
Rayos X
Ionización
Dosimetría de radiaciones
description This project aimed to characterize and calibrate the triple Gas Electron Multiplier (triple-GEM) detector owned by the CRYOMAG research group at Universidad Nacional de Colombia, as well as to determine its possible applications in medical dosimetry. The detector's response to various sources of radiation was characterized in terms of amplitude, duration, shape, and polarity of pulses, as well as its efficiency, gain, energy resolution, and temporal resolution. The triple-GEM detector's energy proportionality was calibrated using the spectrum of an Iron-55 radioactive source. Then, the detector was used to measure radiation doses from this source using four different methods, as well as the dose from a medical portable X-ray machine at varying peak tube voltages and tube currents. Obtained doses were compared to reference values. The characteristic properties of the detector were found to be an energy resolution of 19.5%, a temporal resolution of at least 40ns, and a maximum gain of 5E6. For the dose measurements, a calibration coefficient of 1.13E4 was obtained against the respective reference values of each source of radiation. There was a good agreement between the four dose-measuring methods for the Fe-55 source. It was concluded that the triple-GEM detector can be reliably used as a dosimeter, especially for nuclear medicine applications.
publishDate 2019
dc.date.issued.spa.fl_str_mv 2019-12-17
dc.date.accessioned.spa.fl_str_mv 2020-03-30T06:47:08Z
dc.date.available.spa.fl_str_mv 2020-03-30T06:47:08Z
dc.type.spa.fl_str_mv Trabajo de grado - Maestría
dc.type.driver.spa.fl_str_mv info:eu-repo/semantics/masterThesis
dc.type.version.spa.fl_str_mv info:eu-repo/semantics/acceptedVersion
dc.type.content.spa.fl_str_mv Text
dc.type.redcol.spa.fl_str_mv http://purl.org/redcol/resource_type/TM
status_str acceptedVersion
dc.identifier.uri.none.fl_str_mv https://repositorio.unal.edu.co/handle/unal/77337
dc.identifier.eprints.spa.fl_str_mv http://bdigital.unal.edu.co/75009/
url https://repositorio.unal.edu.co/handle/unal/77337
http://bdigital.unal.edu.co/75009/
dc.language.iso.spa.fl_str_mv spa
language spa
dc.relation.ispartof.spa.fl_str_mv Universidad Nacional de Colombia Sede Bogotá Facultad de Ciencias Departamento de Física
Departamento de Física
dc.relation.haspart.spa.fl_str_mv 53 Física / Physics
dc.relation.references.spa.fl_str_mv Velásquez Moros, Andrea (2019) Characterization and calibration of a triple-GEM detector for medical applications. Maestría thesis, Universidad Nacional de Colombia - Sede Bogotá.
dc.rights.spa.fl_str_mv Derechos reservados - Universidad Nacional de Colombia
dc.rights.coar.fl_str_mv http://purl.org/coar/access_right/c_abf2
dc.rights.license.spa.fl_str_mv Atribución-NoComercial 4.0 Internacional
dc.rights.uri.spa.fl_str_mv http://creativecommons.org/licenses/by-nc/4.0/
dc.rights.accessrights.spa.fl_str_mv info:eu-repo/semantics/openAccess
rights_invalid_str_mv Atribución-NoComercial 4.0 Internacional
Derechos reservados - Universidad Nacional de Colombia
http://creativecommons.org/licenses/by-nc/4.0/
http://purl.org/coar/access_right/c_abf2
eu_rights_str_mv openAccess
dc.format.mimetype.spa.fl_str_mv application/pdf
institution Universidad Nacional de Colombia
bitstream.url.fl_str_mv https://repositorio.unal.edu.co/bitstream/unal/77337/1/1020793274.2019.pdf
https://repositorio.unal.edu.co/bitstream/unal/77337/2/1020793274.2019.pdf.jpg
bitstream.checksum.fl_str_mv 8f859f1dcf28b1dce774face28bbb6ec
cb1b753ebfffb65c213012db58b3c8e7
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
repository.name.fl_str_mv Repositorio Institucional Universidad Nacional de Colombia
repository.mail.fl_str_mv repositorio_nal@unal.edu.co
_version_ 1814089752733810688
spelling Atribución-NoComercial 4.0 InternacionalDerechos reservados - Universidad Nacional de Colombiahttp://creativecommons.org/licenses/by-nc/4.0/info:eu-repo/semantics/openAccesshttp://purl.org/coar/access_right/c_abf2Castro Serrato, Héctor FabioVelásquez Moros, Andrea384a0bc0-8140-4c7d-976d-ad84e59fc8413002020-03-30T06:47:08Z2020-03-30T06:47:08Z2019-12-17https://repositorio.unal.edu.co/handle/unal/77337http://bdigital.unal.edu.co/75009/This project aimed to characterize and calibrate the triple Gas Electron Multiplier (triple-GEM) detector owned by the CRYOMAG research group at Universidad Nacional de Colombia, as well as to determine its possible applications in medical dosimetry. The detector's response to various sources of radiation was characterized in terms of amplitude, duration, shape, and polarity of pulses, as well as its efficiency, gain, energy resolution, and temporal resolution. The triple-GEM detector's energy proportionality was calibrated using the spectrum of an Iron-55 radioactive source. Then, the detector was used to measure radiation doses from this source using four different methods, as well as the dose from a medical portable X-ray machine at varying peak tube voltages and tube currents. Obtained doses were compared to reference values. The characteristic properties of the detector were found to be an energy resolution of 19.5%, a temporal resolution of at least 40ns, and a maximum gain of 5E6. For the dose measurements, a calibration coefficient of 1.13E4 was obtained against the respective reference values of each source of radiation. There was a good agreement between the four dose-measuring methods for the Fe-55 source. It was concluded that the triple-GEM detector can be reliably used as a dosimeter, especially for nuclear medicine applications.Resumen: El objetivo de este proyecto fue caracterizar y calibrar el detector triple Multiplicador de electrones Gaseoso (triple-GEM) que pertenece al grupo de investigación CRYOMAG de la Universidad Nacional de Colombia, al igual que determinar sus posibles aplicaciones en dosimetría médica. La respuesta del detector frente a varias fuentes de radiación fue caracterizada en términos de amplitud, duración, forma y polaridad de los pulsos, como también su eficiencia, ganancia, resolución de energía y resolución temporal. La proporcionalidad de energía del detector triple-GEM fue calibrada utilizando el espectro de una fuente radiactiva de hierro-55. Posteriormente, el detector fue utilizado para medir la dosis de radiación de esta fuente mediante cuatro métodos distintos, como también la dosis de una máquina de rayos X portátil a varios voltajes y corrientes del tubo. Las dosis obtenidas fueron comparadas con los valores de referencia. Las propiedades características del detector fueron una resolución de energía del 19.5%, una resolución temporal de al menos 40ns y una ganancia máxima de 5E6. Para las mediciones de dosis se encontró un coeficiente de calibración de 1.13E4 contra los valores de referencia de cada fuente de radiación. Se dio un buen acuerdo entre los cuatro métodos de medición de dosis que se utilizaron con la fuente de Fe-55. Se concluyó que el detector triple-GEM puede ser utilizado adecuadamente como dosímetro, especialmente para aplicaciones en medicina nuclear.Maestríaapplication/pdfspaUniversidad Nacional de Colombia Sede Bogotá Facultad de Ciencias Departamento de FísicaDepartamento de Física53 Física / PhysicsVelásquez Moros, Andrea (2019) Characterization and calibration of a triple-GEM detector for medical applications. Maestría thesis, Universidad Nacional de Colombia - Sede Bogotá.Characterization and calibration of a triple-GEM detector for medical applicationsTrabajo de grado - Maestríainfo:eu-repo/semantics/masterThesisinfo:eu-repo/semantics/acceptedVersionTexthttp://purl.org/redcol/resource_type/TMParticle detectorsDetector characterizationCosmic radiationRadioactivityX-raysIonizationRadiation dosimetryDetectores de partículasCaracterización de detectoresRadiación cósmicaRadiactividadRayos XIonizaciónDosimetría de radiacionesORIGINAL1020793274.2019.pdfapplication/pdf26066611https://repositorio.unal.edu.co/bitstream/unal/77337/1/1020793274.2019.pdf8f859f1dcf28b1dce774face28bbb6ecMD51THUMBNAIL1020793274.2019.pdf.jpg1020793274.2019.pdf.jpgGenerated Thumbnailimage/jpeg4407https://repositorio.unal.edu.co/bitstream/unal/77337/2/1020793274.2019.pdf.jpgcb1b753ebfffb65c213012db58b3c8e7MD52unal/77337oai:repositorio.unal.edu.co:unal/773372023-07-18 23:03:44.542Repositorio Institucional Universidad Nacional de Colombiarepositorio_nal@unal.edu.co