Evaluation of the dosimetric accuracy of the AAA algorithm (Analytical Anisotropic Algorithm) of the eclipse treatment planning system
Abstract. This current work evaluates the dosimetric accuracy of version 11.0.31 of the AAA algorithm for calculating the photon dose distributions of the Eclipse treatment planning system respect to the input data. The accuracy of the algorithm was analized for the two photon energy spectrums avail...
- Autores:
-
Castellanos Jerez, Armando Cristian Camilo
- Tipo de recurso:
- Fecha de publicación:
- 2016
- Institución:
- Universidad Nacional de Colombia
- Repositorio:
- Universidad Nacional de Colombia
- Idioma:
- spa
- OAI Identifier:
- oai:repositorio.unal.edu.co:unal/59466
- Acceso en línea:
- https://repositorio.unal.edu.co/handle/unal/59466
http://bdigital.unal.edu.co/56973/
- Palabra clave:
- 5 Ciencias naturales y matemáticas / Science
53 Física / Physics
Photon doses distribution
Tissue substituted
Anthropomorphic phantom
Accuracy evaluation of AAA Algorithm
Distribución de dosis de fotones
Tejido substituto
Simulador físico
Evaluación exactitud algoritmo AAA
- Rights
- openAccess
- License
- Atribución-NoComercial 4.0 Internacional
id |
UNACIONAL2_05abaa2e19635be61c19f802387c783f |
---|---|
oai_identifier_str |
oai:repositorio.unal.edu.co:unal/59466 |
network_acronym_str |
UNACIONAL2 |
network_name_str |
Universidad Nacional de Colombia |
repository_id_str |
|
dc.title.spa.fl_str_mv |
Evaluation of the dosimetric accuracy of the AAA algorithm (Analytical Anisotropic Algorithm) of the eclipse treatment planning system |
title |
Evaluation of the dosimetric accuracy of the AAA algorithm (Analytical Anisotropic Algorithm) of the eclipse treatment planning system |
spellingShingle |
Evaluation of the dosimetric accuracy of the AAA algorithm (Analytical Anisotropic Algorithm) of the eclipse treatment planning system 5 Ciencias naturales y matemáticas / Science 53 Física / Physics Photon doses distribution Tissue substituted Anthropomorphic phantom Accuracy evaluation of AAA Algorithm Distribución de dosis de fotones Tejido substituto Simulador físico Evaluación exactitud algoritmo AAA |
title_short |
Evaluation of the dosimetric accuracy of the AAA algorithm (Analytical Anisotropic Algorithm) of the eclipse treatment planning system |
title_full |
Evaluation of the dosimetric accuracy of the AAA algorithm (Analytical Anisotropic Algorithm) of the eclipse treatment planning system |
title_fullStr |
Evaluation of the dosimetric accuracy of the AAA algorithm (Analytical Anisotropic Algorithm) of the eclipse treatment planning system |
title_full_unstemmed |
Evaluation of the dosimetric accuracy of the AAA algorithm (Analytical Anisotropic Algorithm) of the eclipse treatment planning system |
title_sort |
Evaluation of the dosimetric accuracy of the AAA algorithm (Analytical Anisotropic Algorithm) of the eclipse treatment planning system |
dc.creator.fl_str_mv |
Castellanos Jerez, Armando Cristian Camilo |
dc.contributor.advisor.spa.fl_str_mv |
Rozo Albarracín, Edwin (Thesis advisor) |
dc.contributor.author.spa.fl_str_mv |
Castellanos Jerez, Armando Cristian Camilo |
dc.contributor.spa.fl_str_mv |
Plazas de Pinzón, María Cristina |
dc.subject.ddc.spa.fl_str_mv |
5 Ciencias naturales y matemáticas / Science 53 Física / Physics |
topic |
5 Ciencias naturales y matemáticas / Science 53 Física / Physics Photon doses distribution Tissue substituted Anthropomorphic phantom Accuracy evaluation of AAA Algorithm Distribución de dosis de fotones Tejido substituto Simulador físico Evaluación exactitud algoritmo AAA |
dc.subject.proposal.spa.fl_str_mv |
Photon doses distribution Tissue substituted Anthropomorphic phantom Accuracy evaluation of AAA Algorithm Distribución de dosis de fotones Tejido substituto Simulador físico Evaluación exactitud algoritmo AAA |
description |
Abstract. This current work evaluates the dosimetric accuracy of version 11.0.31 of the AAA algorithm for calculating the photon dose distributions of the Eclipse treatment planning system respect to the input data. The accuracy of the algorithm was analized for the two photon energy spectrums available on the linear accelerator Clinac iX: 6 MeV and 15 MeV, was calculated. To perform the evaluation of the AAA algorithm was calculated the percentage differences between its values calculated with the respective measured values taking into account the tolerances and recommendations described in the Task Group report 53. Relative dosimetry was evaluated carefully and it could be observed that a decreasing of this agreement as the depth of the measurement increased, as well as the increase in field size. In the evaluation of the relative dosimetries for both photon energy spectrums, it was found that overall results of percentage agreement was in the inner zone 92.3 % of the points passed the tests, while for the penumbra zone, umbra and build-up the percentage of points which passed the test were 92.4 %, 14.8 % and 100 % respectively. Absolute dosimetry was analyzed in a homogeneous phantom simulator and additionally, we constructed and characterized imageologically a thorax phantom to analyze the accuracy of the algorithm under conditions of inhomogeneity of the medium. |
publishDate |
2016 |
dc.date.issued.spa.fl_str_mv |
2016-05-25 |
dc.date.accessioned.spa.fl_str_mv |
2019-07-02T16:07:41Z |
dc.date.available.spa.fl_str_mv |
2019-07-02T16:07:41Z |
dc.type.spa.fl_str_mv |
Trabajo de grado - Maestría |
dc.type.driver.spa.fl_str_mv |
info:eu-repo/semantics/masterThesis |
dc.type.version.spa.fl_str_mv |
info:eu-repo/semantics/acceptedVersion |
dc.type.content.spa.fl_str_mv |
Text |
dc.type.redcol.spa.fl_str_mv |
http://purl.org/redcol/resource_type/TM |
status_str |
acceptedVersion |
dc.identifier.uri.none.fl_str_mv |
https://repositorio.unal.edu.co/handle/unal/59466 |
dc.identifier.eprints.spa.fl_str_mv |
http://bdigital.unal.edu.co/56973/ |
url |
https://repositorio.unal.edu.co/handle/unal/59466 http://bdigital.unal.edu.co/56973/ |
dc.language.iso.spa.fl_str_mv |
spa |
language |
spa |
dc.relation.ispartof.spa.fl_str_mv |
Universidad Nacional de Colombia Sede Bogotá Facultad de Ciencias Departamento de Física Física Física |
dc.relation.references.spa.fl_str_mv |
Castellanos Jerez, Armando Cristian Camilo (2016) Evaluation of the dosimetric accuracy of the AAA algorithm (Analytical Anisotropic Algorithm) of the eclipse treatment planning system. Maestría thesis, Universidad Nacional de Colombia - Sede Bogotá. |
dc.rights.spa.fl_str_mv |
Derechos reservados - Universidad Nacional de Colombia |
dc.rights.coar.fl_str_mv |
http://purl.org/coar/access_right/c_abf2 |
dc.rights.license.spa.fl_str_mv |
Atribución-NoComercial 4.0 Internacional |
dc.rights.uri.spa.fl_str_mv |
http://creativecommons.org/licenses/by-nc/4.0/ |
dc.rights.accessrights.spa.fl_str_mv |
info:eu-repo/semantics/openAccess |
rights_invalid_str_mv |
Atribución-NoComercial 4.0 Internacional Derechos reservados - Universidad Nacional de Colombia http://creativecommons.org/licenses/by-nc/4.0/ http://purl.org/coar/access_right/c_abf2 |
eu_rights_str_mv |
openAccess |
dc.format.mimetype.spa.fl_str_mv |
application/pdf |
institution |
Universidad Nacional de Colombia |
bitstream.url.fl_str_mv |
https://repositorio.unal.edu.co/bitstream/unal/59466/1/ArmandoC.CastellanosJerez.2016.pdf https://repositorio.unal.edu.co/bitstream/unal/59466/2/ArmandoC.CastellanosJerez.2016.pdf.jpg |
bitstream.checksum.fl_str_mv |
40873d9a9495cfba274ef11c6f2e558e 705961f99fc4e29bd75e883f942f1509 |
bitstream.checksumAlgorithm.fl_str_mv |
MD5 MD5 |
repository.name.fl_str_mv |
Repositorio Institucional Universidad Nacional de Colombia |
repository.mail.fl_str_mv |
repositorio_nal@unal.edu.co |
_version_ |
1814089417159081984 |
spelling |
Atribución-NoComercial 4.0 InternacionalDerechos reservados - Universidad Nacional de Colombiahttp://creativecommons.org/licenses/by-nc/4.0/info:eu-repo/semantics/openAccesshttp://purl.org/coar/access_right/c_abf2Plazas de Pinzón, María CristinaRozo Albarracín, Edwin (Thesis advisor)46576831-b401-4d21-ac8f-8a5d6b6e3a51-1Castellanos Jerez, Armando Cristian Camilo9d1a5e14-15a6-42dc-a3fa-2d2f277d993d3002019-07-02T16:07:41Z2019-07-02T16:07:41Z2016-05-25https://repositorio.unal.edu.co/handle/unal/59466http://bdigital.unal.edu.co/56973/Abstract. This current work evaluates the dosimetric accuracy of version 11.0.31 of the AAA algorithm for calculating the photon dose distributions of the Eclipse treatment planning system respect to the input data. The accuracy of the algorithm was analized for the two photon energy spectrums available on the linear accelerator Clinac iX: 6 MeV and 15 MeV, was calculated. To perform the evaluation of the AAA algorithm was calculated the percentage differences between its values calculated with the respective measured values taking into account the tolerances and recommendations described in the Task Group report 53. Relative dosimetry was evaluated carefully and it could be observed that a decreasing of this agreement as the depth of the measurement increased, as well as the increase in field size. In the evaluation of the relative dosimetries for both photon energy spectrums, it was found that overall results of percentage agreement was in the inner zone 92.3 % of the points passed the tests, while for the penumbra zone, umbra and build-up the percentage of points which passed the test were 92.4 %, 14.8 % and 100 % respectively. Absolute dosimetry was analyzed in a homogeneous phantom simulator and additionally, we constructed and characterized imageologically a thorax phantom to analyze the accuracy of the algorithm under conditions of inhomogeneity of the medium.Resumen En el presente trabajo se eval´ua la exactitud de la versi´on 11.0.31 del algoritmo AAA encargado del c´alculo de las distribuciones de dosis de fotones del sistema de planeaci´on de tratamientos Eclipse respecto a los datos de comisionamiento del acelerador lineal. Se calcul´o la exactitud del algorimo para los dos espectros de energ´ıas de fotones disponibles en el acelerador lineal Clinac iX: de 6 MeV y 15 MeV. Para realizar la evaluaci´on del algoritmo AAA se calcul´o la concordancia entre sus valores calculados con las respectivas mediciones teniendo en cuenta las tolerancias y recomendaciones descritas en el Task Group reporte 53. Se evaluaron dosimetr´ıas relativas en las se pudo ver una disminuci´on en dicha concordancia a medida que aumentaba la profundiad de la medici´on, as´ı mismo con el aumento del tama˜no de campo. En la evaluaci´on de las dosimetr´ıas relativas para ambos espectros de energ´ıa de fotones se obtuvo como resultado que en la zona inner el 92.3 % de los puntos pasaron la prueb, mientras que para las zona de penumbra, umbra y build-up el porcentaje de puntos que pasaron el test fueron 92.4 %, 14.8 % y 100 %, respectivamente. Se analizaron las dosimetr´ıas absolutas en una simulador f´ısico homogeneo y adicionalmente, constru´ımos y caracterizamos imagenologicamente un simulador f´ısico de t´orax para analizar la exactitud del algoritmo en condiciones de inhomogeneidad del medio.Maestríaapplication/pdfspaUniversidad Nacional de Colombia Sede Bogotá Facultad de Ciencias Departamento de Física FísicaFísicaCastellanos Jerez, Armando Cristian Camilo (2016) Evaluation of the dosimetric accuracy of the AAA algorithm (Analytical Anisotropic Algorithm) of the eclipse treatment planning system. Maestría thesis, Universidad Nacional de Colombia - Sede Bogotá.5 Ciencias naturales y matemáticas / Science53 Física / PhysicsPhoton doses distributionTissue substitutedAnthropomorphic phantomAccuracy evaluation of AAA AlgorithmDistribución de dosis de fotonesTejido substitutoSimulador físicoEvaluación exactitud algoritmo AAAEvaluation of the dosimetric accuracy of the AAA algorithm (Analytical Anisotropic Algorithm) of the eclipse treatment planning systemTrabajo de grado - Maestríainfo:eu-repo/semantics/masterThesisinfo:eu-repo/semantics/acceptedVersionTexthttp://purl.org/redcol/resource_type/TMORIGINALArmandoC.CastellanosJerez.2016.pdfapplication/pdf9979417https://repositorio.unal.edu.co/bitstream/unal/59466/1/ArmandoC.CastellanosJerez.2016.pdf40873d9a9495cfba274ef11c6f2e558eMD51THUMBNAILArmandoC.CastellanosJerez.2016.pdf.jpgArmandoC.CastellanosJerez.2016.pdf.jpgGenerated Thumbnailimage/jpeg5074https://repositorio.unal.edu.co/bitstream/unal/59466/2/ArmandoC.CastellanosJerez.2016.pdf.jpg705961f99fc4e29bd75e883f942f1509MD52unal/59466oai:repositorio.unal.edu.co:unal/594662023-04-02 23:04:54.187Repositorio Institucional Universidad Nacional de Colombiarepositorio_nal@unal.edu.co |