Comportamiento estático de un pilar de circona sometido a envejecimiento artificial. Método de elementos finitos

ABSTRACT: Some studies on the effect of zirconia aging mention a degree of reduction of zirconia’s fracture strength varying from 20 to 40%, while other authors argue that aging does not affect the material’s strength. The aim of this study was to evaluate the response of a zirconia abutment subject...

Full description

Autores:
Suárez Blanco, Nini Tatiana
Escobar Restrepo, Julio César
Latorre Correa, Federico
Villarraga Ossa, Junes Abdul
Tipo de recurso:
Article of investigation
Fecha de publicación:
2015
Institución:
Universidad de Antioquia
Repositorio:
Repositorio UdeA
Idioma:
spa
OAI Identifier:
oai:bibliotecadigital.udea.edu.co:10495/4369
Acceso en línea:
http://hdl.handle.net/10495/4369
http://dx.doi.org/10.17533/udea.rfo.v27n1a2
Palabra clave:
Pilar de circona
Envejecimiento circona
Carga estática
Método de elementos finitos
Rights
openAccess
License
Atribución-NoComercial-CompartirIgual 2.5 Colombia (CC BY-NC-SA 2.5 CO)
id UDEA2_74af2825e034f4321f7dfbc8e4b105e8
oai_identifier_str oai:bibliotecadigital.udea.edu.co:10495/4369
network_acronym_str UDEA2
network_name_str Repositorio UdeA
repository_id_str
dc.title.spa.fl_str_mv Comportamiento estático de un pilar de circona sometido a envejecimiento artificial. Método de elementos finitos
dc.title.alternative.spa.fl_str_mv Static behavior of a zirconia abutment subjected to artificial aging. Finite element method
title Comportamiento estático de un pilar de circona sometido a envejecimiento artificial. Método de elementos finitos
spellingShingle Comportamiento estático de un pilar de circona sometido a envejecimiento artificial. Método de elementos finitos
Pilar de circona
Envejecimiento circona
Carga estática
Método de elementos finitos
title_short Comportamiento estático de un pilar de circona sometido a envejecimiento artificial. Método de elementos finitos
title_full Comportamiento estático de un pilar de circona sometido a envejecimiento artificial. Método de elementos finitos
title_fullStr Comportamiento estático de un pilar de circona sometido a envejecimiento artificial. Método de elementos finitos
title_full_unstemmed Comportamiento estático de un pilar de circona sometido a envejecimiento artificial. Método de elementos finitos
title_sort Comportamiento estático de un pilar de circona sometido a envejecimiento artificial. Método de elementos finitos
dc.creator.fl_str_mv Suárez Blanco, Nini Tatiana
Escobar Restrepo, Julio César
Latorre Correa, Federico
Villarraga Ossa, Junes Abdul
dc.contributor.author.none.fl_str_mv Suárez Blanco, Nini Tatiana
Escobar Restrepo, Julio César
Latorre Correa, Federico
Villarraga Ossa, Junes Abdul
dc.subject.none.fl_str_mv Pilar de circona
Envejecimiento circona
Carga estática
Método de elementos finitos
topic Pilar de circona
Envejecimiento circona
Carga estática
Método de elementos finitos
description ABSTRACT: Some studies on the effect of zirconia aging mention a degree of reduction of zirconia’s fracture strength varying from 20 to 40%, while other authors argue that aging does not affect the material’s strength. The aim of this study was to evaluate the response of a zirconia abutment subjected to static loads and artificial aging using the finite element method (FEM). Methods: modeling of the Tapered Screw- Vent implant and the zirconia Zimmer® abutment (Zimmer Dental1 900 Aston Avenue Carlsbad, CA 92008-7308 USA). Four models were designed: one with an implant of 3.7 mm in diameter and a 3.5 mm diameter abutment, another with an implant of 4.7 mm in diameter and a 4.5 mm diameter abutment, and other two with the same dimensions but changing the final fracture limit to 40%, analyzing the response of different components to specific loads. Results: models subjected to decreases in zirconia abutment fracture strength did not show zirconia ifferences in terms of von Mises values. A factor of safety allowed observing the working threshold of the zirconia abutment; failure occurred at values lower than 1. Conclusion: by modifying zirconia’s properties in order to simulate aging, the factor of safety decreases at values lower than 1. However, the applied forces under which the safety factor decreases are higher than normal masticatory forces.
publishDate 2015
dc.date.issued.none.fl_str_mv 2015
dc.date.accessioned.none.fl_str_mv 2016-09-09T20:31:38Z
dc.date.available.none.fl_str_mv 2016-09-09T20:31:38Z
dc.type.spa.fl_str_mv info:eu-repo/semantics/article
dc.type.coarversion.fl_str_mv http://purl.org/coar/version/c_970fb48d4fbd8a85
dc.type.hasversion.spa.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.coar.spa.fl_str_mv http://purl.org/coar/resource_type/c_2df8fbb1
dc.type.redcol.spa.fl_str_mv https://purl.org/redcol/resource_type/ART
dc.type.local.spa.fl_str_mv Artículo de investigación
format http://purl.org/coar/resource_type/c_2df8fbb1
status_str publishedVersion
dc.identifier.citation.spa.fl_str_mv Suárez NT, Escobar JC, Latorre F, Villarraga J. Comportamiento estático de un pilar de circona sometido a envejecimiento artificial. Método de elementos finitos. Rev Fac Odontol Univ Antioq. 2015;27(1): 30-62. http://dx.doi.org/10.17533/udea.rfo.v27n1a2.
dc.identifier.issn.none.fl_str_mv 0121-246X
dc.identifier.uri.none.fl_str_mv http://hdl.handle.net/10495/4369
dc.identifier.doi.none.fl_str_mv http://dx.doi.org/10.17533/udea.rfo.v27n1a2
dc.identifier.eissn.none.fl_str_mv 2145-7670
identifier_str_mv Suárez NT, Escobar JC, Latorre F, Villarraga J. Comportamiento estático de un pilar de circona sometido a envejecimiento artificial. Método de elementos finitos. Rev Fac Odontol Univ Antioq. 2015;27(1): 30-62. http://dx.doi.org/10.17533/udea.rfo.v27n1a2.
0121-246X
2145-7670
url http://hdl.handle.net/10495/4369
http://dx.doi.org/10.17533/udea.rfo.v27n1a2
dc.language.iso.spa.fl_str_mv spa
language spa
dc.relation.ispartofjournalabbrev.spa.fl_str_mv Rev Fac Odontol Univ Antioq
dc.rights.*.fl_str_mv Atribución-NoComercial-CompartirIgual 2.5 Colombia (CC BY-NC-SA 2.5 CO)
dc.rights.spa.fl_str_mv info:eu-repo/semantics/openAccess
dc.rights.uri.*.fl_str_mv https://creativecommons.org/licenses/by-nc-sa/2.5/co/
dc.rights.accessrights.spa.fl_str_mv http://purl.org/coar/access_right/c_abf2
dc.rights.creativecommons.spa.fl_str_mv https://creativecommons.org/licenses/by-nc-sa/4.0/
rights_invalid_str_mv Atribución-NoComercial-CompartirIgual 2.5 Colombia (CC BY-NC-SA 2.5 CO)
https://creativecommons.org/licenses/by-nc-sa/2.5/co/
http://purl.org/coar/access_right/c_abf2
https://creativecommons.org/licenses/by-nc-sa/4.0/
eu_rights_str_mv openAccess
dc.format.mimetype.spa.fl_str_mv application/pdf
dc.publisher.spa.fl_str_mv Universidad de Antioquia, Facultad de Odontología
dc.publisher.place.spa.fl_str_mv Medellín, Colombia
institution Universidad de Antioquia
bitstream.url.fl_str_mv http://bibliotecadigital.udea.edu.co/bitstream/10495/4369/1/SuarezNini_2015_ComportamientoPilarCircona.pdf
http://bibliotecadigital.udea.edu.co/bitstream/10495/4369/5/license.txt
http://bibliotecadigital.udea.edu.co/bitstream/10495/4369/2/license_url
http://bibliotecadigital.udea.edu.co/bitstream/10495/4369/3/license_text
http://bibliotecadigital.udea.edu.co/bitstream/10495/4369/4/license_rdf
bitstream.checksum.fl_str_mv 4f1daf04e51d86249ddf784b37387a3f
8a4605be74aa9ea9d79846c1fba20a33
4afdbb8c545fd630ea7db775da747b2f
d41d8cd98f00b204e9800998ecf8427e
d41d8cd98f00b204e9800998ecf8427e
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
MD5
MD5
MD5
repository.name.fl_str_mv Repositorio Institucional Universidad de Antioquia
repository.mail.fl_str_mv andres.perez@udea.edu.co
_version_ 1812173190273695744
spelling Suárez Blanco, Nini TatianaEscobar Restrepo, Julio CésarLatorre Correa, FedericoVillarraga Ossa, Junes Abdul2016-09-09T20:31:38Z2016-09-09T20:31:38Z2015Suárez NT, Escobar JC, Latorre F, Villarraga J. Comportamiento estático de un pilar de circona sometido a envejecimiento artificial. Método de elementos finitos. Rev Fac Odontol Univ Antioq. 2015;27(1): 30-62. http://dx.doi.org/10.17533/udea.rfo.v27n1a2.0121-246Xhttp://hdl.handle.net/10495/4369http://dx.doi.org/10.17533/udea.rfo.v27n1a22145-7670ABSTRACT: Some studies on the effect of zirconia aging mention a degree of reduction of zirconia’s fracture strength varying from 20 to 40%, while other authors argue that aging does not affect the material’s strength. The aim of this study was to evaluate the response of a zirconia abutment subjected to static loads and artificial aging using the finite element method (FEM). Methods: modeling of the Tapered Screw- Vent implant and the zirconia Zimmer® abutment (Zimmer Dental1 900 Aston Avenue Carlsbad, CA 92008-7308 USA). Four models were designed: one with an implant of 3.7 mm in diameter and a 3.5 mm diameter abutment, another with an implant of 4.7 mm in diameter and a 4.5 mm diameter abutment, and other two with the same dimensions but changing the final fracture limit to 40%, analyzing the response of different components to specific loads. Results: models subjected to decreases in zirconia abutment fracture strength did not show zirconia ifferences in terms of von Mises values. A factor of safety allowed observing the working threshold of the zirconia abutment; failure occurred at values lower than 1. Conclusion: by modifying zirconia’s properties in order to simulate aging, the factor of safety decreases at values lower than 1. However, the applied forces under which the safety factor decreases are higher than normal masticatory forces.RESUMEN: Estudios sobre el efecto del envejecimiento de la circona refieren una disminución de la resistencia a la fractura de la circona que varía del 20 al 40%, mientras que otros argumentan que no influye en la resistencia del material. El propósito de este estudio fue evaluar la respuesta de un pilar de circona sometido a carga estática y envejecimiento artificial usando el método de elementos finitos (MEF). Métodos: se modelaron el implante Tapered Screw-Vent y el pilar de circona Zimmer® (Zimmer Dental1 900 Aston Avenue Carlsbad, CA 92008-7308 USA). Se diseñaron cuatro modelos: uno con implante de 3,7 de diámetro y pilar de 3,5 mm de diámetro, otro con un implante de 4,7 de diámetro y un pilar de 4,5 mm de diámetro, y otros dos con las mismas dimensiones pero modificando el limite último de fractura en un 40%. Se observó el comportamiento de los diferentes componentes ante la carga. Resultados: en los modelos donde se aplicó la disminución de la resistencia a la fractura del pilar de circona, no se observaron diferencias en la circona en cuanto a los valores de von Mises. Se generó un coeficiente de seguridad que permitió observar el umbral de trabajo del pilar de circona, a valores inferiores a 1 se presentó la falla. Conclusión: al modificar las propiedades de la circona, para simular el envejecimiento, el factor de seguridad disminuye a valores inferiores a 1. Sin embargo, las fuerzas aplicadas bajo las cuales disminuye el factor de seguridad son superiores a las fuerzas de la masticación normal.Grupo de Investigación Biomateriales en Odontología, con la participación del Grupo de Diseño Mecánico de la Facultad de Ingeniería, Universidad de Antioquia, Medellín, Colombia.application/pdfspaUniversidad de Antioquia, Facultad de OdontologíaMedellín, Colombiainfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articlehttp://purl.org/coar/resource_type/c_2df8fbb1https://purl.org/redcol/resource_type/ARTArtículo de investigaciónhttp://purl.org/coar/version/c_970fb48d4fbd8a85Atribución-NoComercial-CompartirIgual 2.5 Colombia (CC BY-NC-SA 2.5 CO)info:eu-repo/semantics/openAccesshttps://creativecommons.org/licenses/by-nc-sa/2.5/co/http://purl.org/coar/access_right/c_abf2https://creativecommons.org/licenses/by-nc-sa/4.0/Pilar de circonaEnvejecimiento circonaCarga estáticaMétodo de elementos finitosComportamiento estático de un pilar de circona sometido a envejecimiento artificial. Método de elementos finitosStatic behavior of a zirconia abutment subjected to artificial aging. Finite element methodRev Fac Odontol Univ AntioqRevista Facultad de Odontología Universidad de Antioquia3062271ORIGINALSuarezNini_2015_ComportamientoPilarCircona.pdfSuarezNini_2015_ComportamientoPilarCircona.pdfArtículo de investigaciónapplication/pdf2180787http://bibliotecadigital.udea.edu.co/bitstream/10495/4369/1/SuarezNini_2015_ComportamientoPilarCircona.pdf4f1daf04e51d86249ddf784b37387a3fMD51LICENSElicense.txtlicense.txttext/plain; charset=utf-81748http://bibliotecadigital.udea.edu.co/bitstream/10495/4369/5/license.txt8a4605be74aa9ea9d79846c1fba20a33MD55CC-LICENSElicense_urllicense_urltext/plain; charset=utf-849http://bibliotecadigital.udea.edu.co/bitstream/10495/4369/2/license_url4afdbb8c545fd630ea7db775da747b2fMD52license_textlicense_texttext/html; charset=utf-80http://bibliotecadigital.udea.edu.co/bitstream/10495/4369/3/license_textd41d8cd98f00b204e9800998ecf8427eMD53license_rdflicense_rdfapplication/rdf+xml; charset=utf-80http://bibliotecadigital.udea.edu.co/bitstream/10495/4369/4/license_rdfd41d8cd98f00b204e9800998ecf8427eMD5410495/4369oai:bibliotecadigital.udea.edu.co:10495/43692021-07-30 11:40:19.756Repositorio Institucional Universidad de Antioquiaandres.perez@udea.edu.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