SIMULATION OF A NEONATAL MONITOR FOR MEDICAL TRAINING PURPOSES (SIMULACIÓN DE UN MONITOR NEONATAL PARA ENTRENAMIENTO MÉDICO)
The design of a neonatal monitor for medical training purposes is hereby presented. In order to do that, the following main vital signs were modeled and simulated: ECG, pulse, blood pressure, CO2 level, among others. The signals were integrated to a graphic interface that generates different scenari...
- Autores:
-
Cifuentes, Jenny
Prieto, Flavio
Méndez, Luis Carlos
- Tipo de recurso:
- Article of journal
- Fecha de publicación:
- 2013
- Institución:
- Universidad EIA .
- Repositorio:
- Repositorio EIA .
- Idioma:
- spa
- OAI Identifier:
- oai:repository.eia.edu.co:11190/4804
- Acceso en línea:
- https://repository.eia.edu.co/handle/11190/4804
https://revistas.eia.edu.co/index.php/reveia/article/view/429
- Palabra clave:
- monitor neonatal
vital signs
medical training
signal modeling. Palabras claves
signos vitales
entrenamiento médico
modelado de señales
- Rights
- openAccess
- License
- Revista EIA - 2013
id |
REIA2_d4b2d85dfe6b6628ebc748dd503da6b1 |
---|---|
oai_identifier_str |
oai:repository.eia.edu.co:11190/4804 |
network_acronym_str |
REIA2 |
network_name_str |
Repositorio EIA . |
repository_id_str |
|
dc.title.spa.fl_str_mv |
SIMULATION OF A NEONATAL MONITOR FOR MEDICAL TRAINING PURPOSES (SIMULACIÓN DE UN MONITOR NEONATAL PARA ENTRENAMIENTO MÉDICO) |
dc.title.translated.eng.fl_str_mv |
SIMULATION OF A NEONATAL MONITOR FOR MEDICAL TRAINING PURPOSES (SIMULACIÓN DE UN MONITOR NEONATAL PARA ENTRENAMIENTO MÉDICO) |
title |
SIMULATION OF A NEONATAL MONITOR FOR MEDICAL TRAINING PURPOSES (SIMULACIÓN DE UN MONITOR NEONATAL PARA ENTRENAMIENTO MÉDICO) |
spellingShingle |
SIMULATION OF A NEONATAL MONITOR FOR MEDICAL TRAINING PURPOSES (SIMULACIÓN DE UN MONITOR NEONATAL PARA ENTRENAMIENTO MÉDICO) monitor neonatal vital signs medical training signal modeling. Palabras claves signos vitales entrenamiento médico modelado de señales |
title_short |
SIMULATION OF A NEONATAL MONITOR FOR MEDICAL TRAINING PURPOSES (SIMULACIÓN DE UN MONITOR NEONATAL PARA ENTRENAMIENTO MÉDICO) |
title_full |
SIMULATION OF A NEONATAL MONITOR FOR MEDICAL TRAINING PURPOSES (SIMULACIÓN DE UN MONITOR NEONATAL PARA ENTRENAMIENTO MÉDICO) |
title_fullStr |
SIMULATION OF A NEONATAL MONITOR FOR MEDICAL TRAINING PURPOSES (SIMULACIÓN DE UN MONITOR NEONATAL PARA ENTRENAMIENTO MÉDICO) |
title_full_unstemmed |
SIMULATION OF A NEONATAL MONITOR FOR MEDICAL TRAINING PURPOSES (SIMULACIÓN DE UN MONITOR NEONATAL PARA ENTRENAMIENTO MÉDICO) |
title_sort |
SIMULATION OF A NEONATAL MONITOR FOR MEDICAL TRAINING PURPOSES (SIMULACIÓN DE UN MONITOR NEONATAL PARA ENTRENAMIENTO MÉDICO) |
dc.creator.fl_str_mv |
Cifuentes, Jenny Prieto, Flavio Méndez, Luis Carlos |
dc.contributor.author.spa.fl_str_mv |
Cifuentes, Jenny Prieto, Flavio Méndez, Luis Carlos |
dc.subject.spa.fl_str_mv |
monitor neonatal vital signs medical training signal modeling. Palabras claves signos vitales entrenamiento médico modelado de señales |
topic |
monitor neonatal vital signs medical training signal modeling. Palabras claves signos vitales entrenamiento médico modelado de señales |
description |
The design of a neonatal monitor for medical training purposes is hereby presented. In order to do that, the following main vital signs were modeled and simulated: ECG, pulse, blood pressure, CO2 level, among others. The signals were integrated to a graphic interface that generates different scenarios showing signals of patients with or without pathologies. Simulated signals were validated against real ones and, in general, the error is less than 5 %; in addition, the neonatal monitor was assessed by 16 specialists; those doctors stated that simulated signals are of “excellent quality”, “truthful” and that the interface is “user friendly”.Resumen: Se presenta el desarrollo de un monitor neonatal orientado al entrenamiento médico. Para esto se mo delaron y simularon los principales signos vitales como son: señal ECG, señal de pulso, presión arterial, nivel de CO2, entre otros. Las señales fueron integradas en una interfaz gráfica, la cual permite generar diferentes escenarios de pacientes, no solo normales sino también con patologías. Las señales simuladas fueron validadas contra señales reales y, en general, el error es inferior al 5 %. El monitor neonatal fue evaluado por 16 médicosespecialistas quienes manifestaron que las señales simuladas son “de excelente calidad”, “fidedignas” y que la interfaz es “amigable al usuario”. |
publishDate |
2013 |
dc.date.accessioned.none.fl_str_mv |
2013-11-07 00:00:00 2022-06-17T20:17:15Z |
dc.date.available.none.fl_str_mv |
2013-11-07 00:00:00 2022-06-17T20:17:15Z |
dc.date.issued.none.fl_str_mv |
2013-11-07 |
dc.type.spa.fl_str_mv |
Artículo de revista |
dc.type.eng.fl_str_mv |
Journal article |
dc.type.coar.fl_str_mv |
http://purl.org/coar/resource_type/c_2df8fbb1 |
dc.type.coar.spa.fl_str_mv |
http://purl.org/coar/resource_type/c_6501 http://purl.org/coar/resource_type/c_6501 |
dc.type.driver.spa.fl_str_mv |
info:eu-repo/semantics/article |
dc.type.version.spa.fl_str_mv |
info:eu-repo/semantics/publishedVersion |
dc.type.content.spa.fl_str_mv |
Text |
dc.type.redcol.spa.fl_str_mv |
http://purl.org/redcol/resource_type/ARTREF |
dc.type.coarversion.spa.fl_str_mv |
http://purl.org/coar/version/c_970fb48d4fbd8a85 |
format |
http://purl.org/coar/resource_type/c_6501 |
status_str |
publishedVersion |
dc.identifier.issn.none.fl_str_mv |
1794-1237 |
dc.identifier.uri.none.fl_str_mv |
https://repository.eia.edu.co/handle/11190/4804 |
dc.identifier.eissn.none.fl_str_mv |
2463-0950 |
dc.identifier.url.none.fl_str_mv |
https://revistas.eia.edu.co/index.php/reveia/article/view/429 |
identifier_str_mv |
1794-1237 2463-0950 |
url |
https://repository.eia.edu.co/handle/11190/4804 https://revistas.eia.edu.co/index.php/reveia/article/view/429 |
dc.language.iso.spa.fl_str_mv |
spa |
language |
spa |
dc.relation.bitstream.none.fl_str_mv |
https://revistas.eia.edu.co/index.php/reveia/article/download/429/422 |
dc.relation.citationedition.spa.fl_str_mv |
Núm. 16 , Año 2011 |
dc.relation.citationendpage.none.fl_str_mv |
27 |
dc.relation.citationissue.spa.fl_str_mv |
16 |
dc.relation.citationstartpage.none.fl_str_mv |
9 |
dc.relation.citationvolume.spa.fl_str_mv |
8 |
dc.relation.ispartofjournal.spa.fl_str_mv |
Revista EIA |
dc.rights.spa.fl_str_mv |
Revista EIA - 2013 |
dc.rights.uri.spa.fl_str_mv |
https://creativecommons.org/licenses/by-nc-nd/4.0 |
dc.rights.accessrights.spa.fl_str_mv |
info:eu-repo/semantics/openAccess |
dc.rights.coar.spa.fl_str_mv |
http://purl.org/coar/access_right/c_abf2 |
rights_invalid_str_mv |
Revista EIA - 2013 https://creativecommons.org/licenses/by-nc-nd/4.0 http://purl.org/coar/access_right/c_abf2 |
eu_rights_str_mv |
openAccess |
dc.format.mimetype.spa.fl_str_mv |
application/pdf |
dc.publisher.spa.fl_str_mv |
Fondo Editorial EIA - Universidad EIA |
dc.source.spa.fl_str_mv |
https://revistas.eia.edu.co/index.php/reveia/article/view/429 |
institution |
Universidad EIA . |
bitstream.url.fl_str_mv |
https://repository.eia.edu.co/bitstreams/f59b1162-6ae8-4eb1-8f67-ad863bee255e/download |
bitstream.checksum.fl_str_mv |
790782c1821e038a310d5991d480979b |
bitstream.checksumAlgorithm.fl_str_mv |
MD5 |
repository.name.fl_str_mv |
Repositorio Institucional Universidad EIA |
repository.mail.fl_str_mv |
bdigital@metabiblioteca.com |
_version_ |
1818099038073913344 |
spelling |
Cifuentes, Jennyc1c68df76dbb08783d506c32d1e42469Prieto, Flavio9137d41c5da79263e694d5350fc5af61300Méndez, Luis Carlos33c61682fe692fa647bf7a2fe0d99aa15002013-11-07 00:00:002022-06-17T20:17:15Z2013-11-07 00:00:002022-06-17T20:17:15Z2013-11-071794-1237https://repository.eia.edu.co/handle/11190/48042463-0950https://revistas.eia.edu.co/index.php/reveia/article/view/429The design of a neonatal monitor for medical training purposes is hereby presented. In order to do that, the following main vital signs were modeled and simulated: ECG, pulse, blood pressure, CO2 level, among others. The signals were integrated to a graphic interface that generates different scenarios showing signals of patients with or without pathologies. Simulated signals were validated against real ones and, in general, the error is less than 5 %; in addition, the neonatal monitor was assessed by 16 specialists; those doctors stated that simulated signals are of “excellent quality”, “truthful” and that the interface is “user friendly”.Resumen: Se presenta el desarrollo de un monitor neonatal orientado al entrenamiento médico. Para esto se mo delaron y simularon los principales signos vitales como son: señal ECG, señal de pulso, presión arterial, nivel de CO2, entre otros. Las señales fueron integradas en una interfaz gráfica, la cual permite generar diferentes escenarios de pacientes, no solo normales sino también con patologías. Las señales simuladas fueron validadas contra señales reales y, en general, el error es inferior al 5 %. El monitor neonatal fue evaluado por 16 médicosespecialistas quienes manifestaron que las señales simuladas son “de excelente calidad”, “fidedignas” y que la interfaz es “amigable al usuario”.The design of a neonatal monitor for medical training purposes is hereby presented. In order to do that, the following main vital signs were modeled and simulated: ECG, pulse, blood pressure, CO2 level, among others. The signals were integrated to a graphic interface that generates different scenarios showing signals of patients with or without pathologies. Simulated signals were validated against real ones and, in general, the error is less than 5 %; in addition, the neonatal monitor was assessed by 16 specialists; those doctors stated that simulated signals are of “excellent quality”, “truthful” and that the interface is “user friendly”.Resumen: Se presenta el desarrollo de un monitor neonatal orientado al entrenamiento médico. Para esto se mo delaron y simularon los principales signos vitales como son: señal ECG, señal de pulso, presión arterial, nivel de CO2, entre otros. Las señales fueron integradas en una interfaz gráfica, la cual permite generar diferentes escenarios de pacientes, no solo normales sino también con patologías. Las señales simuladas fueron validadas contra señales reales y, en general, el error es inferior al 5 %. El monitor neonatal fue evaluado por 16 médicosespecialistas quienes manifestaron que las señales simuladas son “de excelente calidad”, “fidedignas” y que la interfaz es “amigable al usuario”.application/pdfspaFondo Editorial EIA - Universidad EIARevista EIA - 2013https://creativecommons.org/licenses/by-nc-nd/4.0info:eu-repo/semantics/openAccessEsta obra está bajo una licencia internacional Creative Commons Atribución-NoComercial-SinDerivadas 4.0.http://purl.org/coar/access_right/c_abf2https://revistas.eia.edu.co/index.php/reveia/article/view/429monitor neonatalvital signsmedical trainingsignal modeling. Palabras clavessignos vitalesentrenamiento médicomodelado de señalesSIMULATION OF A NEONATAL MONITOR FOR MEDICAL TRAINING PURPOSES (SIMULACIÓN DE UN MONITOR NEONATAL PARA ENTRENAMIENTO MÉDICO)SIMULATION OF A NEONATAL MONITOR FOR MEDICAL TRAINING PURPOSES (SIMULACIÓN DE UN MONITOR NEONATAL PARA ENTRENAMIENTO MÉDICO)Artículo de revistaJournal articlehttp://purl.org/coar/resource_type/c_6501http://purl.org/coar/resource_type/c_6501http://purl.org/coar/resource_type/c_2df8fbb1info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionTexthttp://purl.org/redcol/resource_type/ARTREFhttp://purl.org/coar/version/c_970fb48d4fbd8a85https://revistas.eia.edu.co/index.php/reveia/article/download/429/422Núm. 16 , Año 2011271698Revista EIAPublicationOREORE.xmltext/xml2722https://repository.eia.edu.co/bitstreams/f59b1162-6ae8-4eb1-8f67-ad863bee255e/download790782c1821e038a310d5991d480979bMD5111190/4804oai:repository.eia.edu.co:11190/48042023-07-25 16:56:56.228https://creativecommons.org/licenses/by-nc-nd/4.0Revista EIA - 2013metadata.onlyhttps://repository.eia.edu.coRepositorio Institucional Universidad EIAbdigital@metabiblioteca.com |