Agreement between a fitness test and polar s810 for estimating maximum oxygen uptake
Objective Identifying agreement between data for estimating maximum oxygen uptake obtained by a fitness test and the Polar S810 heart rate monitor. Methods This was an observational agreement study comparing two methods for estimating VO2max (simultaneously and independently). Two protocols were app...
- Autores:
- Tipo de recurso:
- Fecha de publicación:
- 2009
- Institución:
- Universidad del Rosario
- Repositorio:
- Repositorio EdocUR - U. Rosario
- Idioma:
- eng
- OAI Identifier:
- oai:repository.urosario.edu.co:10336/23145
- Acceso en línea:
- https://repository.urosario.edu.co/handle/10336/23145
- Palabra clave:
- Exercise
Fitness test
Heart rate
Oxygen consumption
- Rights
- License
- Abierto (Texto Completo)
id |
EDOCUR2_aae1d1127c2d2d8b98868e4253a76ce4 |
---|---|
oai_identifier_str |
oai:repository.urosario.edu.co:10336/23145 |
network_acronym_str |
EDOCUR2 |
network_name_str |
Repositorio EdocUR - U. Rosario |
repository_id_str |
|
spelling |
42aded4b-e9f5-40d0-8b4e-209d9754ac1e-1957ca1f6-0946-4678-9516-794dfcbf669b-1853e74a8-a680-40af-b0f6-41ff3bda0a08-1527164986002020-05-26T00:00:00Z2020-05-26T00:00:00Z2009Objective Identifying agreement between data for estimating maximum oxygen uptake obtained by a fitness test and the Polar S810 heart rate monitor. Methods This was an observational agreement study comparing two methods for estimating VO2max (simultaneously and independently). Two protocols were applied to 46 athletes from the Universidad Nacional de Colombia who had been trained in endurance sports (one at rest and maximum effort i). Results Lin's concordance correlation coefficient CCC) was 0.31 (p and lt;0.003). This coefficient did not change substantially when stratified by gender (male: CCC 0.29; 0.04-0.54 95%CI; female CCC 0.15; 0.14-0.45 95%CI), type of sport (group: CCC 0.38; 0.04-0.54 95%CI; individual: CCC 0.16; 0.20-0.52 95%CI) and its value only improved in athletes aged 24.1 to 28 (CCC 0.65; 0.35-0.93 95%CI). Regarding the maximum test, the Polar S810 gave higher or lower values, with 42 mL/kg/min cut-off. Conclusions Agreement between both methods was generally poor and in the different subgroups of athletes; this meant that the results obtained from estimating VO2max could not be exchanged between the fitness and Polar S810 tests.application/pdf1240064https://repository.urosario.edu.co/handle/10336/23145engUniversidad Nacional de Colombia827No. 5819Revista de Salud PublicaVol. 11Revista de Salud Publica, ISSN:1240064, Vol.11, No.5 (2009); pp. 819-827https://www.scopus.com/inward/record.uri?eid=2-s2.0-77950287568&partnerID=40&md5=68b2b5de5cdf8a275c4d41d2c15f6fd5Abierto (Texto Completo)http://purl.org/coar/access_right/c_abf2instname:Universidad del Rosarioreponame:Repositorio Institucional EdocURExerciseFitness testHeart rateOxygen consumptionAgreement between a fitness test and polar s810 for estimating maximum oxygen uptakeConcordancia en la estimación del consumo máximo de oxígeno entre una prueba de esfuerzo y el polar s810®articleArtículohttp://purl.org/coar/version/c_970fb48d4fbd8a85http://purl.org/coar/resource_type/c_6501Cortés-Reyes E.Echeverry-Raad J.Mancera-Soto E.M.Ramos Caballero, Diana MarcelaORIGINALv11n5a15.pdfapplication/pdf2983741https://repository.urosario.edu.co/bitstreams/abd60aa2-cdb8-4f6d-abc6-997e71fffc7d/download566bf06886f668d434e3d6d06208795cMD51TEXTv11n5a15.pdf.txtv11n5a15.pdf.txtExtracted texttext/plain21508https://repository.urosario.edu.co/bitstreams/347efc92-ab1b-4159-9809-6be23b6a1791/downloada99b53f035fa8bcd4a15ae0b436669cfMD52THUMBNAILv11n5a15.pdf.jpgv11n5a15.pdf.jpgGenerated Thumbnailimage/jpeg3723https://repository.urosario.edu.co/bitstreams/269537bb-5d48-478a-889c-84444ffe1da9/download08aae345839c8c077e30ad0f8a51a406MD5310336/23145oai:repository.urosario.edu.co:10336/231452022-05-02 07:37:20.899697https://repository.urosario.edu.coRepositorio institucional EdocURedocur@urosario.edu.co |
dc.title.spa.fl_str_mv |
Agreement between a fitness test and polar s810 for estimating maximum oxygen uptake |
dc.title.TranslatedTitle.spa.fl_str_mv |
Concordancia en la estimación del consumo máximo de oxígeno entre una prueba de esfuerzo y el polar s810® |
title |
Agreement between a fitness test and polar s810 for estimating maximum oxygen uptake |
spellingShingle |
Agreement between a fitness test and polar s810 for estimating maximum oxygen uptake Exercise Fitness test Heart rate Oxygen consumption |
title_short |
Agreement between a fitness test and polar s810 for estimating maximum oxygen uptake |
title_full |
Agreement between a fitness test and polar s810 for estimating maximum oxygen uptake |
title_fullStr |
Agreement between a fitness test and polar s810 for estimating maximum oxygen uptake |
title_full_unstemmed |
Agreement between a fitness test and polar s810 for estimating maximum oxygen uptake |
title_sort |
Agreement between a fitness test and polar s810 for estimating maximum oxygen uptake |
dc.subject.keyword.spa.fl_str_mv |
Exercise Fitness test Heart rate Oxygen consumption |
topic |
Exercise Fitness test Heart rate Oxygen consumption |
description |
Objective Identifying agreement between data for estimating maximum oxygen uptake obtained by a fitness test and the Polar S810 heart rate monitor. Methods This was an observational agreement study comparing two methods for estimating VO2max (simultaneously and independently). Two protocols were applied to 46 athletes from the Universidad Nacional de Colombia who had been trained in endurance sports (one at rest and maximum effort i). Results Lin's concordance correlation coefficient CCC) was 0.31 (p and lt;0.003). This coefficient did not change substantially when stratified by gender (male: CCC 0.29; 0.04-0.54 95%CI; female CCC 0.15; 0.14-0.45 95%CI), type of sport (group: CCC 0.38; 0.04-0.54 95%CI; individual: CCC 0.16; 0.20-0.52 95%CI) and its value only improved in athletes aged 24.1 to 28 (CCC 0.65; 0.35-0.93 95%CI). Regarding the maximum test, the Polar S810 gave higher or lower values, with 42 mL/kg/min cut-off. Conclusions Agreement between both methods was generally poor and in the different subgroups of athletes; this meant that the results obtained from estimating VO2max could not be exchanged between the fitness and Polar S810 tests. |
publishDate |
2009 |
dc.date.created.spa.fl_str_mv |
2009 |
dc.date.accessioned.none.fl_str_mv |
2020-05-26T00:00:00Z |
dc.date.available.none.fl_str_mv |
2020-05-26T00:00:00Z |
dc.type.eng.fl_str_mv |
article |
dc.type.coarversion.fl_str_mv |
http://purl.org/coar/version/c_970fb48d4fbd8a85 |
dc.type.coar.fl_str_mv |
http://purl.org/coar/resource_type/c_6501 |
dc.type.spa.spa.fl_str_mv |
Artículo |
dc.identifier.issn.none.fl_str_mv |
1240064 |
dc.identifier.uri.none.fl_str_mv |
https://repository.urosario.edu.co/handle/10336/23145 |
identifier_str_mv |
1240064 |
url |
https://repository.urosario.edu.co/handle/10336/23145 |
dc.language.iso.spa.fl_str_mv |
eng |
language |
eng |
dc.relation.citationEndPage.none.fl_str_mv |
827 |
dc.relation.citationIssue.none.fl_str_mv |
No. 5 |
dc.relation.citationStartPage.none.fl_str_mv |
819 |
dc.relation.citationTitle.none.fl_str_mv |
Revista de Salud Publica |
dc.relation.citationVolume.none.fl_str_mv |
Vol. 11 |
dc.relation.ispartof.spa.fl_str_mv |
Revista de Salud Publica, ISSN:1240064, Vol.11, No.5 (2009); pp. 819-827 |
dc.relation.uri.spa.fl_str_mv |
https://www.scopus.com/inward/record.uri?eid=2-s2.0-77950287568&partnerID=40&md5=68b2b5de5cdf8a275c4d41d2c15f6fd5 |
dc.rights.coar.fl_str_mv |
http://purl.org/coar/access_right/c_abf2 |
dc.rights.acceso.spa.fl_str_mv |
Abierto (Texto Completo) |
rights_invalid_str_mv |
Abierto (Texto Completo) http://purl.org/coar/access_right/c_abf2 |
dc.format.mimetype.none.fl_str_mv |
application/pdf |
dc.publisher.spa.fl_str_mv |
Universidad Nacional de Colombia |
institution |
Universidad del Rosario |
dc.source.instname.spa.fl_str_mv |
instname:Universidad del Rosario |
dc.source.reponame.spa.fl_str_mv |
reponame:Repositorio Institucional EdocUR |
bitstream.url.fl_str_mv |
https://repository.urosario.edu.co/bitstreams/abd60aa2-cdb8-4f6d-abc6-997e71fffc7d/download https://repository.urosario.edu.co/bitstreams/347efc92-ab1b-4159-9809-6be23b6a1791/download https://repository.urosario.edu.co/bitstreams/269537bb-5d48-478a-889c-84444ffe1da9/download |
bitstream.checksum.fl_str_mv |
566bf06886f668d434e3d6d06208795c a99b53f035fa8bcd4a15ae0b436669cf 08aae345839c8c077e30ad0f8a51a406 |
bitstream.checksumAlgorithm.fl_str_mv |
MD5 MD5 MD5 |
repository.name.fl_str_mv |
Repositorio institucional EdocUR |
repository.mail.fl_str_mv |
edocur@urosario.edu.co |
_version_ |
1814167619122495488 |